Where have all the ions gone, long time passing? Tandem quadrupole mass spectrometers with atmospheric pressure ionization sensitivity gains since the mid‐1970s. A perspective
Thomas R. Covey
Sciex (Canada)
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
The gains in sensitivity since 1975 for quadrupole mass spectrometers equipped with atmospheric pressure ionization (API), and in particular triple quadrupole mass spectrometers (QqQs) since 1981, have been driven by the needs of the environmental, biomedical, agricultural, and other scientific research, industrial, regulatory, legal, and sporting communities to continually achieve lower limits of quantitation and identification. QqQs have realized a one-million-fold improvement in sensitivity attempting to address these needs over the past two score years. It is the purpose of this article to describe how that came about, not through an exhaustive review of the literature, but rather by describing what general approaches were used across the industry to improve sensitivity and provide some examples to illustrate its evolution. The majority of the gains came from the ion source and its interface to the vacuum system. "Sampling efficiency" is a measurement of the losses in this area so will be a focus of this review. The discovery of the phenomenon of collisional focusing was key to improving sampling efficiency because it enabled designs that increased the ion-containing gas loads from the ion source, using staged differential pumping backed by increasingly larger pumps, and prevented the scattering losses of ions in the resulting gas expansion inside vacuum. Likewise, systems with smaller pumps and lower ion-containing gas loads could be designed with size and cost reduction in mind while maintaining reasonable sampling efficiencies. As a consequence, advancements in the designs of both larger and smaller turbomolecular vacuum pumps were accelerated by pump manufacturers to accommodate the explosive growth in the use of API-QqQ and API-ion trap mass spectrometers that occurred in the 1990s and continued into the new millennium. Sampling efficiency was further improved by increasing the ion yield from electrospray by increasing the rate of droplet desolvation. An estimate of the practical limit to further sensitivity improvements beyond what has been achieved to date is provided to shed light on what to expect in the future. Lastly, the implications and unforeseen consequences of the sensitivity gains are considered with a particular focus on how they have enabled a dramatic increase in daily sample throughput on triple quadrupole and other types of mass spectrometers.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
化学Mass Spectrometry Techniques and Applications
Analytical Chemistry and Chromatography · Pesticide Residue Analysis and Safety
参考文献 132
此处列出前 3 条
引用本文 17
按被引量排序,此处列出前 3 条