Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
Felix K. Wegner, Simon Kochhäuser, Gerrit Frommeyer, Philipp S. Lange, Christian Ellermann, Patrick Leitz, Patrick Müller, Julia Köbe 等 10 位
University Hospital Münster West German Heart and Vascular Center Essen Marienhospital Osnabrück
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
INTRODUCTION: Supraventricular tachycardias (SVT) are often difficult to document due to their intermittent, short-lasting nature. Smartphone-based one-lead ECG monitors (sECG) were initially developed for the diagnosis of atrial fibrillation. No data have been published regarding their potential role in differentiating inappropiate sinus tachycardia (IST) from regular SVT. If cardiologists could distinguish IST from SVT in sECG, economic health care burden might be significantly reduced. METHODS: We prospectively recruited 75 consecutive patients with known SVT undergoing an EP study. In all patients, four ECG were recorded: a sECG during SVT and during sinus tachycardia and respective 12-lead ECG. Two experienced electrophysiologists were blinded to the diagnoses and separately evaluated all ECG. RESULTS: , 60 AVNRT, 15 AVRT). The electrophysiologists' blinded interpretation of sECG recordings showed a sensitivity of 89% and a specificity of 91% for the detection of SVT (interobserver agreement κ = 0.76). In high-quality sECG recordings (68%), sensitivity rose to 95% with a specificity of 92% (interobserver agreement of κ = 0.91). Specificity increased to 96% when both electrophysiologists agreed on the diagnosis. Respective 12-lead ECG had a sensitivity of 100% and specificity of 98% for the detection of SVT. CONCLUSION: A smartphone-based one-lead ECG monitor allows for differentiation of SVT from IST in about 90% of cases. These results should encourage cardiologists to integrate wearables into clinical practice, possibly reducing time to definitive diagnosis of an arrhythmia and unnecessary EP procedures. A smartphone-based one lead ECG device (panel A) can be used reliably to differentiate supraventricular tachycardia (panel B) from inappropriate sinus tachycardia when compared to a simultaneously conducted gold-standard electrophysiology study (panels C, D).
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Cardiovascular Syncope and Autonomic Disorders
Cardiac Arrhythmias and Treatments · Atrial Fibrillation Management and Outcomes
参考文献 14
此处列出前 3 条
引用本文 17
按被引量排序,此处列出前 3 条