A non-enzymatic microneedle electrochemical biosensor based on Pt@PDA-MWCNTs modified electrode for in vivo and in situ glucose detection
Yan Li, Jiamin Wu, Hao Zhou, Shurong Shen, Yunxiang Zhu, Lei Nie, Guohua Jiang
Zhejiang Sci-Tech University Hangzhou Normal University Wenzhou Medical University First Affiliated Hospital of Wenzhou Medical University
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Real-time glucose monitoring is essential for diabetes care, but existing non-enzymatic electrochemical biosensors have great challenges, including poor selectivity, insufficient stability and bad long-term reliability. In this work, a non-enzymatic electrochemical biosensor integrated with a porous microneedle array and a Pt@PDA-MWCNTs electrode was developed for continuous in vivo and in situ glucose monitoring. The microneedle component had a built-in reservoir and iontophoresis system. The developed biosensor exhibited a sensitive glucose response within the linear range of 0.1–30 mM, with a sensitivity of 3.91 μA·mM −1 ·cm −2 , a detection limit of 0.14 mM, and favorable selectivity against common interferents, including ascorbic acid, dopamine, uric acid, chloride ions, and bovine serum albumin (BSA). In addition, the sensor enables stable long‑term operation, resists biofouling, and retains its performance even in complex biological matrices. In diabetes rat models, the biosensor showed high consistency with commercial glucometers over short-term (4.5-h) and long-term (12-h) monitoring, with Mean Absolute Relative Difference (MARD) values below 5% and all data points in zone A of the Clarke error grid analysis (CEGA). Histological examination, cytotoxicity and hemolysis tests showed no obvious inflammation, apoptosis or tissue damage, proving good biocompatibility and rapid skin recovery after insertion. This strategy provides a feasible approach for the non-enzymatic continuous glucose monitoring, and holds promising potential for long-term diabetes management.
逐年被引趋势
暂无年度引用数据
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
工程Electrochemical sensors and biosensors
Electrochemical Analysis and Applications · Advanced biosensing and bioanalysis techniques
参考文献 71
此处列出前 3 条