TEM-EDS microanalysis: Comparison among the standardless, Cliff & Lorimer and absorption correction quantification methods
Roberto Conconi, Gennaro Ventruti, Fernando Nieto, Giancarlo Capitani
University of Milano-Bicocca University of Bari Aldo Moro Universidad de Granada
内容与影响
Available quantification methods for energy dispersive X-ray microanalysis in transmission electron microscopy, such as the standardless method (SLM), the Cliff-Lorimer approximation (CLA) and the absorption correction method (ACM), are compared. As expected, the CLA and ACM give superior results with respect to the SLM. As far as absorption can be considered negligible, CLA and ACM perform similarly. However, starting from mass-thickness of the order of 22×10−6 g/cm2, absorption become significant and the ACM gives better results. More accurate analyses can be obtained with the ACM if distinct kO/Si factors are determined for light and heavy minerals, respectively, placing a divide at 2.90 g/cm3. Caution must be used when k-factors are derived indirectly from minerals with very different structure/chemistry, suggesting that separate k-factors data sets are required for accurate EDS quantification, at least for the major and diverse broad classes of minerals. Element diffusion of monovalent cations and channelling effects may represent a complication, especially in very anisotropic minerals such as phyllosilicates, where these two phenomena may occur together.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
回答优先基于摘要、文献信息与可获取全文;依据不足时会明确说明。
学术脉络
学科主题
材料 / 化学Electron and X-Ray Spectroscopy Techniques
X-ray Spectroscopy and Fluorescence Analysis · Surface and Thin Film Phenomena
参考文献 28
此处列出前 3 条
施引文献 20
按被引量排序,此处列出前 3 条