CE-Microreactor-CE-MS/MS for Protein Analysis
Regine M. Schoenherr, Mingliang Ye, Michael W. Vannatta, Norman J. Dovic̀hi
University of Washington
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
We present a proof-of-principle for a fully automated bottom-up approach to protein characterization. Proteins are first separated by capillary electrophoresis. A pepsin microreactor is incorporated into the distal end of this capillary. Peptides formed in the reactor are transferred to a second capillary, where they are separated by capillary electrophoresis and characterized by mass spectrometry. While peptides generated from one digestion are being separated in the second capillary, the next protein fraction undergoes digestion in the microreactor. The migration time in the first dimension capillary is characteristic of the protein while migration time in the second dimension is characteristic of the peptide. Spot capacity for the two-dimensional separation is 590. A MS/MS analysis of a mixture of cytochrome c and myoglobin generated Mascot MOWSE scores of 107 for cytochrome c and 58 for myoglobin. The sequence coverages were 48% and 22%, respectively.