Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children
Muhterem Duyu, Anıl Doğan Bektaş, Zeynep Karakaya, Meral Bahar, Aybüke Gunalp, Yasemin Mocan Çağlar, Meryem Nihal Yersel, Özlem Bozkurt
Istanbul Medeniyet University
内容与影响
The objective of this study was to evaluate a novel microstream method by comparison with PaCO 2 and the more standard mainstream capnometer in intubated pediatric patients. We hypothesized that the novel microstream method would superior compared to the traditional mainstream method in predicting PaCO 2 . This was a prospective single-center comparative study. The study was carried out on 174 subjects with a total of 1338 values for each method. Data were collected prospectively from mainstream and microstream capnometer simultaneously and compared with PaCO 2 results. Although both mainstream PetCO 2 (mainPetCO 2 ) and microstream PetCO 2 (microPetCO 2 ) were moderately correlated (r = 0.63 and r = 0.68, respectively) with PaCO 2 values, mainPetCO 2 was in better agreement with PaCO 2 in all subjects (bias ± precision values of 3.8 ± 8.9 and 7.3 ± 8.2 mmHg, respectively). In those with severe pulmonary disease, the mainPetCO 2 and microPetCO 2 methods were highly correlated with PaCO 2 (r = 0.80 and r = 0.81, respectively); however, the biases of both methods increased (14.8 ± 9.1 mmHg and 16.2 ± 9.0 mmHg, respectively). In cases with increased physiologic dead space ventilation, the agreement levels of mainPetCO 2 and microPetCO 2 methods became distorted (bias ± precision values of 20.9 ± 11.2 and 25.0 ± 11.8 mm Hg, respectively) even though mainPetCO 2 and microPetCO 2 were highly correlated (r = 0.78 and r = 0.78, respectively). It was found that the novel microstream capnometer method for PetCO 2 measurements provided no superiority to the traditional mainstream method. Both capnometer methods may be useful in predicting the trend of PaCO 2 due to significant correlations with the gold standard measurement in cases with severe pulmonary disease or increased physiological dead space –despite reduced accuracy.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
回答优先基于摘要、文献信息与可获取全文;依据不足时会明确说明。
学术脉络
学科主题
生物医学Respiratory Support and Mechanisms
Airway Management and Intubation Techniques · Cardiac Arrest and Resuscitation
参考文献 39
此处列出前 3 条
施引文献 9
按被引量排序,此处列出前 3 条