Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy
Shuang Xu, Yuanming Hu, Boyuan Li, Christina R. Inscoe, Donald A. Tyndall, Yueh Z. Lee, Jianping Lü, Otto Z. Zhou
University of North Carolina at Chapel Hill
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Cone beam computed tomography (CBCT) is widely used in medical and dental imaging. Compared to a multidetector CT, it provides volumetric images with high isotropic resolution at a reduced radiation dose, cost and footprint without the need for patient translation. The current CBCT has several intrinsic limitations including reduced soft tissue contrast, inaccurate quantification of X-ray attenuation, image distortions and artefacts, which have limited its clinical applications primarily to imaging hard tissues and made quantitative analysis challenging. Here we report a multisource CBCT (ms-CBCT) which overcomes the short-comings of the conventional CBCT by using multiple narrowly collimated and rapidly scanning X-ray beams from a carbon nanotube field emission source array. Phantom imaging studies show that, the ms-CBCT increases the accuracy of the Hounsfield unit values by 60%, eliminates the cone beam artefacts, extends the axial coverage, and improves the soft tissue contrast-to-noise ratio by 30-50%, compared to the CBCT configuration.