Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy
Shuang Xu, Yuanming Hu, Boyuan Li, Christina R. Inscoe, Donald A. Tyndall, Yueh Z. Lee, Jianping Lü, Otto Zhou
University of North Carolina at Chapel Hill Applied Physical Sciences (United States)
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Cone beam computed tomography (CBCT) is widely used in medical and dental imaging. Compared to a multidetector CT, it provides volumetric images with high isotropic resolution at a reduced radiation dose, cost and footprint without the need for patient translation. The current CBCT has several intrinsic limitations including reduced soft tissue contrast, inaccurate quantification of X-ray attenuation, image distortions and artefacts, which have limited its clinical applications primarily to imaging hard tissues and made quantitative analysis challenging. Here we report a multisource CBCT (ms-CBCT) which overcomes the short-comings of the conventional CBCT by using multiple narrowly collimated and rapidly scanning X-ray beams from a carbon nanotube field emission source array. Phantom imaging studies show that, the ms-CBCT increases the accuracy of the Hounsfield unit values by 60%, eliminates the cone beam artefacts, extends the axial coverage, and improves the soft tissue contrast-to-noise ratio by 30-50%, compared to the CBCT configuration.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Medical Imaging Techniques and Applications
Advanced X-ray and CT Imaging · Dental Radiography and Imaging
参考文献 40
此处列出前 3 条
引用本文 14
按被引量排序,此处列出前 3 条