An automatic beam alignment system based on relative reference points for Thomson scattering diagnosis system
Wei-qiang Tan, Peng Yuan, Peng Wang, Wenzhe Mao, Jinlin Xie, Tao Lan, Weixing Ding, Hong Li 等 11 位
University of Science and Technology of China Shanghai Jiao Tong University
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
An automatic beam alignment system based on relative reference points is developed for the Thomson scattering system on Keda Torus eXperiment. Two critical apertures around the vacuum vessel are designed to shield stray light, and a probe beam is required to go through the centers of these two apertures, which are the reference points for alignment. Since these two apertures are coated with light absorbing materials, three fibers with glowing tips are employed to indicate the centers of two apertures. CMOS cameras are used to monitor beam deviations. The misalignment correction is achieved by driving piezomotor mirror mounts via a program developed with LabVIEW, which includes the image acquisition and processing module and the deviation correction module. As a result, this system can correct beam misalignment in less than 20 s and suppress the long-term drift of laser pointing in ±10 μrad. Also, this system has the capability to correct up to about 2.3 mm of camera shift with our experiment condition.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
物理Magnetic confinement fusion research
Laser-Plasma Interactions and Diagnostics · Particle accelerators and beam dynamics
参考文献 5
此处列出前 3 条
引用本文 1
按被引量排序,此处列出前 3 条