Design parameter investigation of a cold-cathode Penning ion source for general laboratory applications
Joshua L. Rovey
Missouri University of Science and Technology Starfire Industries (United States)
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
An investigation of different design parameters of a cold-cathode Penning ion source is described. The Penning source had an iron cathode body, samarium cobalt permanent magnet, stainless steel anode and iron cathode faceplate. The faceplate extraction orifice diameter, anode shape and permanent magnet location were adjusted. Discharge current–voltage characteristics and current to a grounded target were measured for the different ion source configurations. Results indicate that for hydrogen and helium operation, the largest orifice, cylinder anode and baseline magnet location provides optimum source operation. Specifically, a continuous positive hydrogen ion target current of 1.0 ± 0.15 mA was attainable at only 1.6 × 10 −4 Torr of pressure. At 1.0 mTorr of hydrogen pressure a target current of 2.0 ± 0.15 mA was possible. These results far surpass the design goal of 1 mA hydrogen ion beam at 1 mTorr of pressure. This cold-cathode Penning ion source design has potential application in a wide variety of laboratory research and development projects.