Development of cryogenic cooling system using a double G-M cryo-cooler for NMR spectrometer
Hideki Tanaka, Yosuke Fukuda, Morihito Okada, Norihide Saho, K. Saitoh, Hitoshi Kitaguchi
Hitachi (Japan) National Institute for Materials Science
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
The sensitivity of nuclear magnetic resonance (NMR) spectrometer is limited by thermal noise. To increase the signal-noise-ratio (SNR), cryogenically cooled probes with cryo-coolers have been used. In order to cool RF coils in the probe below 10K, Gifford- McMahon/Joule-Thomson (G-M/J-T) cryo-coolers are used. But, the G-M/J-T cryo-coolers aren't suitable for a long time operation. On the other hand, G-M cryo-coolers have high reliability. We designed a cryogenically cooled probe system using a double G-M cryo-cooler with an RF coil and preamplifiers that could be cooled below 5 and 50K, respectively. This probe system consists of a probe unit, a transfer unit and a cryo-cooling unit. The cryo-cooling unit consists of two G-M cryo-coolers, two counter-flow heat exchangers, helium gas lines, etc.,. Compressed helium gas is cooled by the cooling stages of G-M cryo-coolers and two heat exchangers. The cooled helium gas is supplied to the probe unit through the 2.5m long transfer unit which consists of a 45mm diameter flexible vacuum vessel containing four helium gas lines. Cold stages for the RF coil and the preamplifiers in the probe unit are cooled by the cryogenically cooled helium gas. Experimental results showed that the cold stage temperature for the RF coil and for preamplifiers after 12 hours was 5.29 and 40.3K, respectively.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
化学Advanced NMR Techniques and Applications
NMR spectroscopy and applications · Atomic and Subatomic Physics Research
参考文献 2
此处列出前 3 条
引用本文 7
按被引量排序,此处列出前 3 条