Technical Feasibility and Performance Studies of a 0.009-inch Nitinol Microguidewire for Potential Neuroendovascular Applications
Christopher M. Putman, John C. Chaloupka, Purushothaman Kailasnath, Jonathan Alderman
Yale University
内容与影响
RATIONALE AND OBJECTIVES: The authors conduct technical feasibility and performance studies on a new 0.009-inch nitinol microguidewire for potential neuroendovascular applications. METHODS: In vivo microcatheterizations of brachiocephalic arteries were performed in four pigs using the 0.009-inch nitinol microguidewire with a commonly used microcatheter and compared with the performance of commercially available microguidewires. Vessels catheterized using the 0.009-inch wire were evaluated by histopathological analysis. A standardized, bench-top method of evaluating basic mechanical properties of microguidewires also was used to compare the 0.009-inch nitinol microguidewire with two commonly used microguidewires. RESULTS: The 0.009-inch microguidewire had similar steerability, tractability, torque control, and distal tip flexibility to the commonly used microguidewires in the in vivo simulations. Bench top testing showed the 0.009-inch microguidewire to have comparable distal tip flexibility and objective mechanical properties to the commonly used microguidewires. CONCLUSIONS: The comparable subjective and objective mechanical properties of the 0.009-inch nitinol microguidewire to that of commonly used microguidewires further establishes the possibility of clinical implementation.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
回答优先基于摘要、文献信息与可获取全文;依据不足时会明确说明。
学术脉络
学科主题
生物医学Peripheral Nerve Disorders
Spine and Intervertebral Disc Pathology · Pain Management and Treatment
参考文献 3
此处列出前 3 条
施引文献 3
按被引量排序,此处列出前 3 条