Plane wave medical ultrasound imaging using adaptive beamforming
Iben Kraglund Holfort, Fredrik Gran, Jørgen Arendt Jensen
Technical University of Denmark GN Store Nord (Denmark)
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
In this paper, the adaptive, minimum variance (MV) beam-former is applied to medical ultrasound imaging. The significant resolution and contrast gain provided by the adaptive, minimum variance (MV) beamformer, introduces the possibility of plane wave (PW) ultrasound imaging. Data is obtained using Field II and a 7 MHz, 128-elements, linear array transducer with lambda/2-spacing. MV is compared to the conventional delay-and-sum (DS) beamformer with Boxcar and Hanning weights. Furthermore, the PW images are compared to the a conventional ultrasound image, obtained from a linear scan sequence. The four approaches, {Linear Scan, DS Boxcar, DS Hanning, MV}, have full width at half maximum of {0.82, 0.71, 1.28, 0.12} mm and peak side-lobe levels of {-40.1,-16.8,-34.4,-57.0} dB.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Ultrasound Imaging and Elastography
Ultrasonics and Acoustic Wave Propagation · Photoacoustic and Ultrasonic Imaging
参考文献 11
此处列出前 3 条
引用本文 28
按被引量排序,此处列出前 3 条