Plane wave medical ultrasound imaging using adaptive beamforming
Iben Kraglund Holfort, Fredrik Gran, Jørgen Arendt Jensen
Technical University of Denmark GN Store Nord (Denmark)
内容与影响
In this paper, the adaptive, minimum variance (MV) beam-former is applied to medical ultrasound imaging. The significant resolution and contrast gain provided by the adaptive, minimum variance (MV) beamformer, introduces the possibility of plane wave (PW) ultrasound imaging. Data is obtained using Field II and a 7 MHz, 128-elements, linear array transducer with lambda/2-spacing. MV is compared to the conventional delay-and-sum (DS) beamformer with Boxcar and Hanning weights. Furthermore, the PW images are compared to the a conventional ultrasound image, obtained from a linear scan sequence. The four approaches, {Linear Scan, DS Boxcar, DS Hanning, MV}, have full width at half maximum of {0.82, 0.71, 1.28, 0.12} mm and peak side-lobe levels of {-40.1,-16.8,-34.4,-57.0} dB.