Extending Reach Inside the MRI Bore: A 7-DOF, Low-Friction, Hydrostatic Teleoperator
Samuel Frishman, Robert D. Ings, Vipul R. Sheth, Bruce Lewis Daniel, Mark R. Cutkosky
Stanford University Stanford Medicine
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
We present a hydrostatic teleoperator that provides physicians remote access inside the MRI bore and enables real-time MRI guided interventions such as liver biopsies. The device consists of a custom six-axis arm and a needle insertion end effector. The manipulator is passive and backdrivable with a near one-to-one mapping of motions and forces between the input and output. The six-axis arm translates and orients the needle during insertion and passively reflects respiratory motion while maintaining contact with the skin surface. Arm joints employ novel rotary rolling-diaphragms that provide stiff and low-friction rotational motion without the need for cables, belts, or gear mechanisms found in other solutions. We perform experiments to characterize the device's force and position tracking and demonstrate its functionality with path following tasks. Our results find a system roll-off frequency at 20Hz and that teleoperation tasks are performed comparably to holding the output directly. We motivate the need for force transparency in MRI guided biopsies with a user study in which five radiologists perform phantom membrane puncture biopsies using the needle insertion mechanism. The results indicate a 50% reduction in scans and a 14.5% decrease in membrane overshoot when force feedback is present.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
工程Teleoperation and Haptic Systems
Soft Robotics and Applications · Robotic Mechanisms and Dynamics
参考文献 37
此处列出前 3 条
引用本文 11
按被引量排序,此处列出前 3 条