Doped Plastic Scintillator Properties for Soft Tissue Dosimetry
Suffian Mohamad Tajudin, Yoshihito Namito, Toshiya Sanami, Hideo Hirayama, Suhairul Hashim
Sultan Zainal Abidin University High Energy Accelerator Research Organization University of Technology Malaysia
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
In medical dosimetry, the radiation quantity of interest is frequently the dose to tissue. To make direct dosimeters that show the dose absorbed in soft tissue, tissue-equivalent material is preferred. Plastic scintillators (PLSs) are made of materials with a density similar to soft tissue and relatively have energy independent response. Non-proportionality must be taken into account when reproducing the light output spectrum for photons below 200 keV, as it reduces light output significantly. In comparison to a standard PLS, both low and high-energy photons’ pulse height distribution spectra as a result of lead doping were measured and calculated. The effect of a 0.5% lead doping scintillator is sufficient to make up for the loss or difference between soft tissues and PLSs in response to low-energy photons, particularly below 200 keV.