Orbitrap™-SIMS for accurate quantification of arsenic in SiGe across blanket and fin-type geometries
Rita Tilmann, Alexis Franquet, A. A. Merkulov, Lennaert Wouters, Clément Porret, Alexander Pirkl, Paul van der Heide
IMEC
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Accurate quantification of n-type dopant, such as phosphorus (P) and arsenic (As) in silicon-germanium (SiGe), is critical for source/drain engineering, junction control, and device variability in advanced complementary metal-oxide-semiconductor. Especially for As in SiGe, conventional microanalysis techniques, such as transmission electron microscopy combined with energy dispersive x-ray spectroscopy or electron energy loss spectroscopy, struggle at technologically relevant doping levels and suffer from peak-overlap issues for As in the presence of Ge. Common secondary ion mass spectrometry (SIMS) platforms, such as magnetic sector (MS), quadrupole (Quad), and time-of-flight (ToF), face severe mass interferences at limited mass resolving power as well as low ion yields. Here, we translate the Orbitrap™-SIMS workflow, previously validated for P in Si/SiGe, to As in SiGe, combine it with self-focusing SIMS (SF-SIMS) for patterned features, and benchmark Orbitrap™-SIMS against ToF-SIMS, Quad-SIMS, and MS-SIMS. Using blanket Si1−xGex standards and patterned Ge1−y−zSiySnz:As fins, we show that the exceptional mass resolving power of the Orbitrap™ mass analyzer eliminates the dominant As interferences (e.g., As− vs GeH−; AsSi− vs GeSiH−/29SiGe−/30Si73Ge−; and AsGe− vs 73Ge76Ge−/Ge2H−) and detection limits improve by >1 order of magnitude versus ToF-SIMS and MS-SIMS when SF-relevant clusters are used. As a result, only Orbitrap™-SIMS recovers the physically correct “top-only” As-doped segment in Ge1−y−zSiySnz fins, whereas other mass analyzers erroneously report As throughout the fin due to unresolved overlaps and high backgrounds. These results are confirmed by fast Fourier transform-scanning spreading resistance microscopy 2D carrier distribution mapping.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
工程Ion-surface interactions and analysis
Mass Spectrometry Techniques and Applications · Semiconductor materials and devices
参考文献 21
此处列出前 3 条
引用本文 1
按被引量排序,此处列出前 3 条