False elevation of serum cortisol in chemiluminescence immunoassay by Siemens Advia Centaur XP system in 21-hydroxylase deficiency: an ‘endocrine laboma’
Neeti Agrawal, Partha Pratim Chakraborty, Anirban Sinha, Animesh Maiti
Medical College and Hospital, Kolkata
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Liquid chromatography-mass spectrometry, the gold standard method for cortisol measurement, is expensive and not widely available in the developing countries. Chemiluminescent immunoassay, commonly used for cortisol measurement is prone to clinically meaningful inter-assay variability in some analysers. This occurs due to non-specific nature of anticortisol antibodies used in different platforms, having cross reactivity with structurally similar cortisol precursors like 17α-hydroxyprogesterone (17OHP), 11-deoxycortisol and 21-deoxycortisol. In patients with 21-hydroxylase deficiency, where 17OHP and 21-deoxycortisol are significantly elevated, older generation machines like Siemens Advia Centaur XP provide spuriously high cortisol concentration compared with values measured by Roche Cobas e 411 or Siemens Immulite 1000. Diagnosis of potentially life-threatening salt-wasting 21-hydroxylase deficiency may be missed and treatment may be delayed due to such interference. Two children with classic 21-hydroxylase deficiency are being reported here, in whom high cortisol values were observed in Siemens Advia Centaur XP system.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Sexual Differentiation and Disorders
Growth Hormone and Insulin-like Growth Factors · Adrenal Hormones and Disorders
参考文献 10
此处列出前 3 条
引用本文 7
按被引量排序,此处列出前 3 条