Determination of frequency response from step response: application to fluid-filled catheters
Stanton A. Glantz, John V. Tyberg
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
The performance of a fluid-filled catheter can be described by reporting its undamped natural frequency and damping ratio. These parameters can be measured by subjecting the catheter to sinusoidally varying pressures at a wide variety of frequencies to obtain the frequency response. They can also be computed from the response to a step change in pressure, which is often easier to produce. This paper derives the required equations and includes a graph which permits one to look up the undamped natural frequency after measuring the period and decay rate of the oscillation following a step change in pressure.