Predicting post-operative vault and optimal implantable collamer lens size using machine learning based on various ophthalmic device combinations
Xi Chen, Yiming Ye, Huan Yao, Chang Liu, Anqi He, Xiangtao Hou, Keming Zhao, Zedu Cui 等 14 位
Sun Yat-sen University
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
BACKGROUND: Implantable Collamer Lens (ICL) surgery has been proven to be a safe, effective, and predictable method for correcting myopia and myopic astigmatism. However, predicting the vault and ideal ICL size remains technically challenging. Despite the growing use of artificial intelligence (AI) in ophthalmology, no AI studies have provided available choices of different instruments and combinations for further vault and size predictions. This study aimed to fill this gap and predict post-operative vault and appropriate ICL size utilizing the comparison of numerous AI algorithms, stacking ensemble learning, and data from various ophthalmic devices and combinations. RESULTS: = 0.499 (95% CI 0.470-0.528), mean absolute error = 130.655 (95% CI 128.949-132.111), accuracy = 0.895 (95% CI 0.883-0.907), AUC = 0.928 (95% CI 0.916-0.941)]. Sulcus-to-sulcus (STS), a parameter from UBM, ranked among the top five significant contributors to both post-operative vault and optimal ICL size prediction, consistently outperforming white-to-white (WTW). Moreover, dual-device combinations or single-device parameters could also effectively predict vault and ideal ICL size, and excellent ICL selection prediction was achievable using only UBM parameters. CONCLUSIONS: Strategies based on multiple machine learning algorithms for different ophthalmic devices and combinations are applicable for vault predicting and ICL sizing, potentially improving the safety of the ICL implantation. Moreover, our findings emphasize the crucial role of UBM in the perioperative period of ICL surgery, as it provides key STS measurements that outperformed WTW measurements in predicting post-operative vault and optimal ICL size, highlighting its potential to enhance ICL implantation safety and accuracy.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Corneal surgery and disorders
Ophthalmology and Visual Impairment Studies · Intraocular Surgery and Lenses
参考文献 35
此处列出前 3 条
引用本文 28
按被引量排序,此处列出前 3 条