A Comparative Evaluation of Air-Jet Sieve for Particle Size Analysis of Pharmaceutical Powders
Shabir Z. Masih, J. H. Shukla, K. S. Rajagopalan
Massachusetts College of Pharmacy and Health Sciences
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
The performance of the air-jet sieve (AJS), for the particle size analysis of pharmaceutical powder, was evaluated and compared with the conventional sieve shaker (ESS). Two grades of acetaminophen (APAP) were used as test pharmaceutical powders. The particle size analysis by the AJS was simple, reproducible, efficient and accurate. The problematic powder, possessing excessive electrostatic charges and ranging in mean diameter from 17 nm to 800 nm (beyond the scope of ESS) could be evaluated accurately by AJS. The analysis time required by AJS was relatively less than that required by ESS. The accuracy of the determination by AJS was not affected by the sample size. Both the speed of rotation (vibrations and/or shaking) and the total time of sifting affected the results of the determination by ESS. The particle size value of APAP-special determined by ESS and AJS were 680.0 ± 0.14 nm and 40.0 ± 0.32 respectively, whereas the value assigned by the manufacturer was below 74 nm (-200 mesh). The precision of either method for particle size analysis of powder possessing excessive electrostatic charge was improved by the use of carbon black as an anti-static agent.