Interference of neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators: An in vitro study
Salome Ryf, Thomas Wölber, Fırat Duru, Roger Luechinger
ETH Zurich Institute for Biomedical Engineering Federal Office of Public Health University Hospital Zurich
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Permanent magnets may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets have become widely available in recent years and are incorporated in various articles of daily life. We conducted an in-vitro study to evaluate the ability of NdFeB magnets for home and office use to cause interference with cardiac pacemakers and ICDs. The magnetic fields of ten NdFeB magnets of different size and shape were measured at increasing distances beginning from the surface until a field-strength (B-field) value of 0.5 mT was reached. Furthermore, for each magnet the distance was determined at which a sample pacemaker switched from magnet mode to normal mode. Depending on the size and remanence of individual magnets, a B-field value of 0.5 mT was found at distances ranging from 1.5 cm to 30 cm and a value of 1 mT at distances from 1 cm to 22 cm. The pacemaker behavior was influenced at distances from 1 cm to 24 cm. NdFeB magnets for home and office use may cause interference with cardiac pacemakers and ICDs at distances up to 24 centimeters. Patient education and product declarations should include information about the risk associated with these magnets.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
生物医学Cardiac pacing and defibrillation studies
Orthopaedic implants and arthroplasty · Magnetic Properties of Alloys
参考文献 0
引用本文 35
按被引量排序,此处列出前 3 条