Accuracy Improvement of a Compact 85Rb Atom Gravimeter by Suppressing Laser Crosstalk and Light Shift
G.-G. Ge, Xi Chen, Jinting Li, Danfang Zhang, Meng He, Wenzhang Wang, Yang Zhou, Jiaqi Zhong 等 12 位
Chinese Academy of Sciences University of Chinese Academy of Sciences Hefei University
阅读操作
确认中在文库中上传 PDF 后可生成中文音频讲解。
摘要与影响
Rb atom gravimeter (AG). The diameter of the sensor head is 35 cm and the height is 65 cm; the optical and electronic systems are installed in four standard 3U cabinets. The measurement accuracy of this AG is improved by suppress laser crosstalk and light shift. In addition, the angle of the Raman laser reflector is adjusted and locked, and the attitude of the sensing head is automatically adjusted, and the vibration noise is also compensated. The comparison measurement results between this AG and the superconducting gravimeter indicate that its long-term stability is 0.65 μGal @50000 s.
逐年被引趋势
关键指标
同类平均 = 1
同领域 · 同年份 · 同类型
Google Scholar 与 OpenAlex 的被引统计范围不同,数值存在差异属正常。
AI 辅助阅读
依据:摘要
可就本文提问;依据不足时会说明。
学术脉络
学科主题
工程Inertial Sensor and Navigation
Geophysics and Sensor Technology · Advanced Frequency and Time Standards
参考文献 31
此处列出前 3 条
引用本文 13